ASL H6

© 2006 by EST · Last updated: 29.09.2006

ASL Model H6 is a complete eye tracking system for use in situations where the subject can wear lightweight, head mounted optics and must have unrestricted freedom of movement.  It includes all necessary equipment to begin work immediately.  The system control unit is compact in size.  The optics are lightweight and mounted on an adjustable headband.  The scene is recorded with a color camera that can be mounted on the headband or on a fixed tripod.  The advantages to a fixed scene camera (non-head mounted) include a better, more stable image, less to wear on the head and less to set up and adjust.  The images from the eye and scene cameras are displayed on two external 9” monitors.  ASL EYEPOS operating software and EYENAL off-line data analysis software programs are provided for installation on a PC or laptop computer.

Data Output

The Model H6 is designed to measure a subject’s eye line of gaze with respect to the head.  The measurement is displayed as a cursor or set of cross hairs  superimposed on the image from the scene camera.  A videotape of this image can be created as a permanent record and used for data analysis.

Recorded data include time, x and y eye position coordinates, and pupil diameter.  External data events/marks can be recorded along with the eye tracker data.  If the Model H6 includes the optional magnetic head tracker, head position and orientation are also recorded.  Up to 20 individual surfaces of interest can be defined in the EYEHEAD environment.

Model H6 EYEPOS operating software provides the system operator with the ability to enter calibration and subject data, and specify the operating parameters of the Model H6.  EYEPOS also converts the eye tracker data records into ASCII format for transmission to other computers or for off-line spreadsheet analysis.  Data is available directly from the Model 6000 control unit through a serial port (RS232).  The Model H6’s interface PC can be connected to an Ethernet network, permitting data analysis from remote locations.   ASL will supply a dll and any additional required information for integration.

Data Analysis Software

EYENAL data analysis software is a set of off-line analysis programs for displaying and processing eye position and pupil diameter data that have been recorded with the Model H6.  EYENAL programs identify fixations, plot scan patterns, let the user define areas of interest on the stimulus scene, tabulate pupil diameter and compute various statistical parameters.

SPECIFICATIONS  

Head Mounted Optics

  • Sampling and Output Rates

    • 60 Hz (NTSC) or 50 Hz (PAL)

    • 120, 240 and 360 Hz (optional High Speed)

  • Measurement principle:  pupil-corneal reflection

  • System accuracy:  0.5 degree visual angle

  • Resolution:  0.1 degree visual angle

  • Head movement:  unlimited

  • Visual range:  45 degrees horizontal, 35 degrees vertical

  • Weight 8 oz (includes headband, optics module monocle and scene camera

Included equipment

  • Series 6000 Control Unit

  • Headband Mounted Optics

  • Head Mounted Scene Camera (color)

  • Black and White Display Monitors (x2), LCD Display Monitors (x2), or two PCI framgrabbers

  • EYEPOS operating software

  • EYENAL data analysis software

Additional Options: 

EYEHEAD Integration - Hardware and software for integrating eye and head position data

Download the H6 Brochure

Also see: EyeHead Integration (EHI)

All proprietary names and product names are trademarks or registered trademarks of their respective owners. EST doesn´t offer any guarantee for the technical details of the manufacturers.